TY - JOUR
T1 - Voxel-based dose reconstruction for total body irradiation with helical tomotherapy
AU - Chao, Ming
AU - Penagaricano, Jose
AU - Yan, Yulong
AU - Moros, Eduardo G.
AU - Corry, Peter
AU - Ratanatharathorn, Vaneerat
PY - 2012/4/1
Y1 - 2012/4/1
N2 - Purpose: We have developed a megavoltage CT (MVCT)-based dose reconstruction strategy for total body irradiation (TBI) with helical TomoTherapy (HT) using a deformable registration model to account for the patient's interfraction changes. The proposed technique serves as an efficient tool for delivered dose verification and, potentially, plan adaptation. Methods and Materials: Four patients with acute myelogenous leukemia treated with TBI using HT were selected for this study. The prescription was 12 Gy, 2 Gy/fraction, twice per day, given at least 6 h apart. The original plan achieved coverage of 80% of the clinical target volume (CTV) by the 12 Gy isodose surface. MVCTs were acquired prior to each treatment. Regions of interest were contoured on each MVCT. The dose for each fraction was calculated based on the MVCT using the HT planned adaptive station. B-spline deformable registration was conducted to establish voxel-to-voxel correspondence between the MVCT and the planning CT. The resultant deformation vector was employed to map the reconstructed dose from each fraction to the same point as the plan dose, and a voxel-to-voxel summed dose from all six fractions was obtained. The reconstructed dose distribution and its dosimetric parameters were compared with those of the original treatment plan. Results: While changes in CTV contours occurred in all patients, the reconstructed dose distribution showed that the dose-volume histogram for CTV coverage was close (<1.5%) to that of the original plan. For sensitive structures, the differences between the reconstructed and the planned doses were less than 3.0%. Conclusion: Voxel-based dose reconstruction strategy that takes into account interfraction anatomical changes using MVCTs is a powerful tool for treatment verification of the delivered doses. This proposed technique can also be applied to adaptive TBI therapy using HT.
AB - Purpose: We have developed a megavoltage CT (MVCT)-based dose reconstruction strategy for total body irradiation (TBI) with helical TomoTherapy (HT) using a deformable registration model to account for the patient's interfraction changes. The proposed technique serves as an efficient tool for delivered dose verification and, potentially, plan adaptation. Methods and Materials: Four patients with acute myelogenous leukemia treated with TBI using HT were selected for this study. The prescription was 12 Gy, 2 Gy/fraction, twice per day, given at least 6 h apart. The original plan achieved coverage of 80% of the clinical target volume (CTV) by the 12 Gy isodose surface. MVCTs were acquired prior to each treatment. Regions of interest were contoured on each MVCT. The dose for each fraction was calculated based on the MVCT using the HT planned adaptive station. B-spline deformable registration was conducted to establish voxel-to-voxel correspondence between the MVCT and the planning CT. The resultant deformation vector was employed to map the reconstructed dose from each fraction to the same point as the plan dose, and a voxel-to-voxel summed dose from all six fractions was obtained. The reconstructed dose distribution and its dosimetric parameters were compared with those of the original treatment plan. Results: While changes in CTV contours occurred in all patients, the reconstructed dose distribution showed that the dose-volume histogram for CTV coverage was close (<1.5%) to that of the original plan. For sensitive structures, the differences between the reconstructed and the planned doses were less than 3.0%. Conclusion: Voxel-based dose reconstruction strategy that takes into account interfraction anatomical changes using MVCTs is a powerful tool for treatment verification of the delivered doses. This proposed technique can also be applied to adaptive TBI therapy using HT.
KW - Deformable registration
KW - Dose reconstruction
KW - Helical tomotherapy
KW - Total body irradiation
UR - http://www.scopus.com/inward/record.url?scp=84858699717&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2011.01.021
DO - 10.1016/j.ijrobp.2011.01.021
M3 - Article
C2 - 21470791
AN - SCOPUS:84858699717
SN - 0360-3016
VL - 82
SP - 1575
EP - 1583
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 5
ER -