TY - JOUR
T1 - Validation of whole room indirect calorimeters
T2 - Refinement of current methodologies
AU - Rising, Russell
AU - Foerster, Thomas
AU - Arad, Avigdor D.
AU - Albu, Jeanine
AU - Pi-Sunyer, Xavier
N1 - Funding Information:
This study was supported in part by a National Institutes of Health Obesity Center Core grant (no. P30-DK026687).
Publisher Copyright:
© 2017 Columbia University.
PY - 2017/11
Y1 - 2017/11
N2 - Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO2) and nitrogen (N2) by a precision blender (PB). To determine the best method, PC of 6, 10, 22-h and PB infusions of 6, 10, and 14-h, were conducted. The 14-h infusion consisted of two metabolic settings. Energy expenditure (EE; kJ), ventilation (V; liters/min) of oxygen (VO2), VCO2, and respiratory quotient (VCO2/VO2) obtained from the WRIC were extrapolated to the respective test durations and compared to similarly calculated values. Moreover, accurate equations (AE) were derived to correct infusions for additional N2. As a final evaluation of a PC validated WRIC, weight maintenance (WM), energy balance (EB), respiratory quotient (RQ), and food quotients (FQ) were determined in 22 subjects who had repeat 24-h EE measurements. Statistical analyses (P < 0.05) were conducted (SPSS, version 23). Significant differences in RQ existed between PC and stoichiometry after 6-h. Errors for the rest of the PC tests ranged from –1.5 ± 2.4 (VCO2) to 2.8 ± 4.6% (EE). When compared with the WRIC, all uncorrected metabolic parameters for six and 10-h PB infusions were significantly different with errors from –12.8 ± 1.6 (VO2) to 6.0 ± 2.8% (RQ). The AE reduced the magnitude of errors to –12.4 ± 1.5 (RQ) to 2.2 ± 3.0% (RQ). The PB infusion with two settings showed similar performance. No differences in WM, EB, RQ, or FQ existed in the subjects. In conclusion, 10-h PC tests are sufficient for validating WRICs.
AB - Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO2) and nitrogen (N2) by a precision blender (PB). To determine the best method, PC of 6, 10, 22-h and PB infusions of 6, 10, and 14-h, were conducted. The 14-h infusion consisted of two metabolic settings. Energy expenditure (EE; kJ), ventilation (V; liters/min) of oxygen (VO2), VCO2, and respiratory quotient (VCO2/VO2) obtained from the WRIC were extrapolated to the respective test durations and compared to similarly calculated values. Moreover, accurate equations (AE) were derived to correct infusions for additional N2. As a final evaluation of a PC validated WRIC, weight maintenance (WM), energy balance (EB), respiratory quotient (RQ), and food quotients (FQ) were determined in 22 subjects who had repeat 24-h EE measurements. Statistical analyses (P < 0.05) were conducted (SPSS, version 23). Significant differences in RQ existed between PC and stoichiometry after 6-h. Errors for the rest of the PC tests ranged from –1.5 ± 2.4 (VCO2) to 2.8 ± 4.6% (EE). When compared with the WRIC, all uncorrected metabolic parameters for six and 10-h PB infusions were significantly different with errors from –12.8 ± 1.6 (VO2) to 6.0 ± 2.8% (RQ). The AE reduced the magnitude of errors to –12.4 ± 1.5 (RQ) to 2.2 ± 3.0% (RQ). The PB infusion with two settings showed similar performance. No differences in WM, EB, RQ, or FQ existed in the subjects. In conclusion, 10-h PC tests are sufficient for validating WRICs.
KW - Blender
KW - Energy
KW - Infusion and combustion
KW - Propane
KW - Validation
UR - http://www.scopus.com/inward/record.url?scp=85036558061&partnerID=8YFLogxK
U2 - 10.14814/phy2.13521
DO - 10.14814/phy2.13521
M3 - Article
C2 - 29180485