Abstract
Background: The Omnipod® 5 Automated Insulin Delivery (AID) System was shown to be safe and effective following 3 months of use in people with type 1 diabetes (T1D); however, data on the durability of these results are limited. This study evaluated the long-term safety and effectiveness of Omnipod 5 use in people with T1D during up to 2 years of use. Materials and Methods: After a 3-month single-arm, multicenter, pivotal trial in children (6–13.9 years) and adolescents/adults (14–70 years), participants could continue system use in an extension phase. HbA1c was measured every 3 months for up to 15 months; continuous glucose monitor metrics were collected for up to 2 years. Results: Participants (N = 224) completed median (interquartile range) 22.3 (21.7, 22.7) months of AID. HbA1c was reduced in the pivotal trial from 7.7% – 0.9% in children and 7.2% – 0.9% in adolescents/adults to 7.0% – 0.6% and 6.8% – 0.7%, respectively, (P < 0.0001), and was maintained at 7.2% – 0.7% and 6.9% – 0.6% after 15 months (P < 0.0001 from baseline). Time in target range (70–180 mg/dL) increased from 52.4% – 15.6% in children and 63.6% – 16.5% in adolescents/adults at baseline to 67.9% – 8.0% and 73.8% – 10.8%, respectively, during the pivotal trial (P < 0.0001) and was maintained at 65.9% – 8.9% and 72.9% – 11.3% during the extension (P < 0.0001 from baseline). One episode of diabetic ketoacidosis and seven episodes of severe hypoglycemia occurred during the extension. Children and adolescents/adults spent median 96.1% and 96.3% of time in Automated Mode, respectively. Conclusion: Our study supports that long-term use of the Omnipod 5 AID System can safely maintain improvements in glycemic outcomes for up to 2 years of use in people with T1D.
Original language | English |
---|---|
Pages (from-to) | 11-23 |
Number of pages | 13 |
Journal | Diabetes Technology and Therapeutics |
Volume | 26 |
Issue number | 1 |
DOIs | |
State | Published - 1 Jan 2024 |
Keywords
- Artificial pancreas
- Clinical trials
- Closed-loop systems
- Insulin pumps
- Type 1 diabetes