Triggered electromyography for placement of thoracic pedicle screws: Is it reliable?

  • Amer F. Samdani
  • , Mark Tantorski
  • , Patrick J. Cahill
  • , Ashish Ranade
  • , Stephen Koch
  • , David H. Clements
  • , Randal R. Betz
  • , Jahangir Asghar

Research output: Contribution to journalReview articlepeer-review

35 Scopus citations

Abstract

Reliable electromyography (EMG) thresholds for detecting medial breaches in the thoracic spine are lacking, and there is a paucity of reports evaluating this modality in patients with adolescent idiopathic scoliosis (AIS). This retrospective analysis evaluates the ability of triggered EMG to detect medial breaches with thoracic pedicle screws in patients with AIS. We reviewed 50 patients (937 pedicle screws) undergoing posterior spinal fusion (PSF) with intraoperative EMG testing. Postoperative CT scans were used for breach identification, and EMG values were analyzed. There were 47 medial breaches noted with a mean threshold stimulus of 10.2 mA (milliamperes). Only 8/47 breaches stimulated at 2-6 mA. Thirteen of the forty-seven screws tested at an EMG value ≤6 mA and/or a decrease of ≥65% compared with intraosseously placed screws. The sensitivity and positive predictive value for EMG was 0.28 and 0.21. A subanalysis of T10-T12 screws identified six of seven medial breaches. Using guidelines from the current literature, EMG does not appear to be reliable in detecting medial breaches from T2 to T9 but may have some utility from T10 to T12.

Original languageEnglish
Pages (from-to)869-874
Number of pages6
JournalEuropean Spine Journal
Volume20
Issue number6
DOIs
StatePublished - Jun 2011
Externally publishedYes

Keywords

  • Adolescent idiopathic scoliosis
  • Pedicle screws
  • Posterior spinal fusion
  • Thoracic
  • Triggered electromyography

Fingerprint

Dive into the research topics of 'Triggered electromyography for placement of thoracic pedicle screws: Is it reliable?'. Together they form a unique fingerprint.

Cite this