TY - JOUR
T1 - Total aortic arch replacement using a frozen elephant trunk device
T2 - Results of a 1-year US multicenter trial
AU - Coselli, Joseph S.
AU - Roselli, Eric E.
AU - Preventza, Ourania
AU - Malaisrie, S. Chris
AU - Stewart, Allan
AU - Stelzer, Paul
AU - Takayama, Hiroo
AU - Chen, Edward P.
AU - Estrera, Anthony L.
AU - Gleason, Thomas G.
AU - Fischbein, Michael P.
AU - Girardi, Leonard N.
AU - Patel, Himanshu J.
AU - Bavaria, Joseph E.
AU - LeMaire, Scott A.
N1 - Publisher Copyright:
© 2022 The American Association for Thoracic Surgery
PY - 2024/5
Y1 - 2024/5
N2 - Objective: In this prospective US investigational device exemption trial, we assessed the safety and 1-year clinical outcomes of the Thoraflex Hybrid device (Terumo Aortic) for the frozen elephant trunk technique to repair the ascending aorta, aortic arch, and descending thoracic aorta. Methods: For the trial, which involved 12 US sites, 65 patients without rupture were recruited into the primary study group, and 9 patients were recruited into the rupture group. All patients underwent open surgical repair of the ascending aorta, aortic arch, and descending thoracic aorta in cases of aneurysm and/or dissection. The primary end point was freedom from major adverse events (MAE), defined as permanent stroke, permanent paraplegia/paraparesis, unanticipated aortic-related reoperation (excluding reoperation for bleeding), or all-cause mortality. Results: In the primary study group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 81% (51/63). Seven patients (11%) died (including 2 before 30 days or discharge), 3 patients (5%) suffered permanent stroke, and 3 (5%) developed permanent paraplegia/paraparesis. Twenty-six patients (41%) underwent planned extension procedures, including 22 endovascular procedures within a median of 122 (interquartile range, 64-156) days. In the aortic rupture group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 71% (5/7). One patient (14%) died, 2 patients (29%) had permanent stroke, and none had permanent paraplegia/paraparesis. No extension procedures were performed in the rupture group. Conclusions: One-year results with the Thoraflex Hybrid device are acceptable. Long-term data are necessary to assess the durability of these repairs.
AB - Objective: In this prospective US investigational device exemption trial, we assessed the safety and 1-year clinical outcomes of the Thoraflex Hybrid device (Terumo Aortic) for the frozen elephant trunk technique to repair the ascending aorta, aortic arch, and descending thoracic aorta. Methods: For the trial, which involved 12 US sites, 65 patients without rupture were recruited into the primary study group, and 9 patients were recruited into the rupture group. All patients underwent open surgical repair of the ascending aorta, aortic arch, and descending thoracic aorta in cases of aneurysm and/or dissection. The primary end point was freedom from major adverse events (MAE), defined as permanent stroke, permanent paraplegia/paraparesis, unanticipated aortic-related reoperation (excluding reoperation for bleeding), or all-cause mortality. Results: In the primary study group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 81% (51/63). Seven patients (11%) died (including 2 before 30 days or discharge), 3 patients (5%) suffered permanent stroke, and 3 (5%) developed permanent paraplegia/paraparesis. Twenty-six patients (41%) underwent planned extension procedures, including 22 endovascular procedures within a median of 122 (interquartile range, 64-156) days. In the aortic rupture group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 71% (5/7). One patient (14%) died, 2 patients (29%) had permanent stroke, and none had permanent paraplegia/paraparesis. No extension procedures were performed in the rupture group. Conclusions: One-year results with the Thoraflex Hybrid device are acceptable. Long-term data are necessary to assess the durability of these repairs.
KW - aortic aneurysm
KW - aortic arch (total)
KW - aortic dissection
KW - frozen elephant trunk
KW - hybrid repair
UR - http://www.scopus.com/inward/record.url?scp=85189836974&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2022.08.029
DO - 10.1016/j.jtcvs.2022.08.029
M3 - Article
C2 - 36253292
AN - SCOPUS:85189836974
SN - 0022-5223
VL - 167
SP - 1680-1692.e2
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
IS - 5
ER -