TY - JOUR
T1 - Three to four years outcomes of the absorb bioresorbable vascular scaffold versus second-generation drug-eluting stent
T2 - A meta-analysis
AU - Goel, Sunny
AU - Pasam, Ravi Teja
AU - Chava, Srilekha
AU - Sharma, Abhishek
AU - Malik, Bilal Ahmad
AU - Ayzenberg, Sergey
AU - Frankel, Robert
AU - Shani, Jacob
AU - Gidwani, Umesh
N1 - Publisher Copyright:
© 2019 Wiley Periodicals, Inc.
PY - 2020/2/1
Y1 - 2020/2/1
N2 - Objective: This meta-analysis sought to evaluate the outcomes of absorb bioresorbable vascular scaffolds (BVS) compared with second-generation drug-eluting stents (DES) after 3 years, the approximate time of complete polymer bioresorption. Background: BVS were found to be inferior to second-generation DES in early and mid-term outcomes with a higher rate of target vessel myocardial infarction (TV-MI) and device thrombosis (DT). Improper implantation techniques and incomplete bioresorption of the poly-l-lactide (PLLA) polymer were sighted as possible reasons. Methods: We conducted an electronic database search for all randomized control trials that compared absorb BVS to second-generation DES and reported outcomes of interest after 3 years of absorb BVS implantation. Assuming interstudy heterogeneity, a random-effects analysis was conducted with odds ratio as the effect size of choice to compare the event rates between the two groups. Results: A total of four studies (n = 3,245, BVS = 2075, DES = 1,170) were included in the final analysis. Pooled analysis revealed that there was no difference between absorb BVS and second-generation DES with respect to target lesion failure (TLF) (OR = 1.23, 95% CI = 0.73–2.07, p = 0.44), TV-MI (OR = 1.03, 95% CI = 0.42–2.53, p = 0.95), target lesion revascularization (TLR) (OR = 1.61, 95% CI = 0.77–3.33, p = 0.20) and definite/probable DT (OR = 0.71, 95% CI = 0.10–5.07, p = 0.74). Also, there was no difference in cardiac mortality (OR = 0.66, 95% CI = 0.22–1.94, p = 0.45). Conclusions: Between 3 and 4 years of follow-up, patients receiving absorb BVS did not have significantly different outcomes, in terms of TLF, TV-MI, TLR, DT, and cardiac mortality, compared to DES.
AB - Objective: This meta-analysis sought to evaluate the outcomes of absorb bioresorbable vascular scaffolds (BVS) compared with second-generation drug-eluting stents (DES) after 3 years, the approximate time of complete polymer bioresorption. Background: BVS were found to be inferior to second-generation DES in early and mid-term outcomes with a higher rate of target vessel myocardial infarction (TV-MI) and device thrombosis (DT). Improper implantation techniques and incomplete bioresorption of the poly-l-lactide (PLLA) polymer were sighted as possible reasons. Methods: We conducted an electronic database search for all randomized control trials that compared absorb BVS to second-generation DES and reported outcomes of interest after 3 years of absorb BVS implantation. Assuming interstudy heterogeneity, a random-effects analysis was conducted with odds ratio as the effect size of choice to compare the event rates between the two groups. Results: A total of four studies (n = 3,245, BVS = 2075, DES = 1,170) were included in the final analysis. Pooled analysis revealed that there was no difference between absorb BVS and second-generation DES with respect to target lesion failure (TLF) (OR = 1.23, 95% CI = 0.73–2.07, p = 0.44), TV-MI (OR = 1.03, 95% CI = 0.42–2.53, p = 0.95), target lesion revascularization (TLR) (OR = 1.61, 95% CI = 0.77–3.33, p = 0.20) and definite/probable DT (OR = 0.71, 95% CI = 0.10–5.07, p = 0.74). Also, there was no difference in cardiac mortality (OR = 0.66, 95% CI = 0.22–1.94, p = 0.45). Conclusions: Between 3 and 4 years of follow-up, patients receiving absorb BVS did not have significantly different outcomes, in terms of TLF, TV-MI, TLR, DT, and cardiac mortality, compared to DES.
KW - absorb
KW - bioresorbable vascular scaffolds
KW - clinical outcomes
KW - drug-eluting stents
UR - http://www.scopus.com/inward/record.url?scp=85064693355&partnerID=8YFLogxK
U2 - 10.1002/ccd.28290
DO - 10.1002/ccd.28290
M3 - Article
C2 - 31002216
AN - SCOPUS:85064693355
SN - 1522-1946
VL - 95
SP - 216
EP - 223
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
IS - 2
ER -