Skip to main navigation Skip to search Skip to main content

Seven to twelve year results with Versys ET cementless stem. A retrospective study of 225 cases

  • Fabio D'Angelo
  • , Luigi Murena
  • , Ettore Vulcano
  • , Giovanni Zatti
  • , Paolo Cherubino

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

A retrospective review was conducted to evaluate the mid-term results (8.5 years follow-up) of the Versys ET stem (Zimmer, Warsaw, Indiana, USA). From 1995 to 2000, 225 total hip arthroplasties (THA) were performed using this device. All patients were evaluated clinically by the Harris Hip Score, and radiographically. The Hip Score increased from 54 preoperatively to 97 points postoperatively. The stem displayed a varus alignment between 5° and 10° in 17 cases without any clinical consequences (HHS 96.8). We recorded one case of septic loosening and one case of aseptic loosening. The stability of the stem was radiographically evaluated according to Engh's criteria, confirming bone ongrowth in all other cases.

Original languageEnglish
Pages (from-to)81-86
Number of pages6
JournalHIP International
Volume20
Issue number1
DOIs
StatePublished - 2010
Externally publishedYes

Keywords

  • Cementless prosthesis
  • Hip arthroplasty
  • Straight stem
  • Titanium
  • Varus alignment

Fingerprint

Dive into the research topics of 'Seven to twelve year results with Versys ET cementless stem. A retrospective study of 225 cases'. Together they form a unique fingerprint.