TY - JOUR
T1 - Relative contribution of trabecular and cortical bone to primary implant stability
T2 - An in vitro model study
AU - Wang, Russell
AU - Eppell, Steven J.
AU - Nguyen, Christian
AU - Morris, Nathan
PY - 2016/4
Y1 - 2016/4
N2 - The specific aim of this study was to examine the relative contributions to the implant insertion torque value (ITV) by cortical and trabecular components of an in vitro bone model. Simulated bone blocks of polyurethane were used with 2 densities of foam (0.08 g/cm3 to mimic trabecular bone and 0.64 g/cm3 to mimic cortical bone). We have developed a new platform technology to collect data that enables quantitative evaluation of ITV at different implant locations. Seven groups were used to model varying thicknesses of cortical bone over a lower-quality trabecular bone that have clinical significance: A solid 0.08 g/cm3 block; 1 mm, 2 mm, and 3 mm thick 0.64 g/cm3 sheets with no underlayer; and 1 mm, 2 mm, and 3 mm thick 0.64 g/cm3 sheets laminated on top of a 4 cm thick 0.08 g/cm3 block. The ITVs were recorded as a function of insertion displacement distance. Relative contributions of ITV ranged from 3% to 18% from trabecular bone, and 62% to 74% from cortical bone depending on the thickness of the cortical layer. Inserting an implant into 2-mm and 3-mm cortical layers laminated atop trabecular blocks had a synergistic effect on ITVs. Finally, an implant with a reverse bevel design near the abutment showed final average torque values that were 14% to 34% less than their maximum torque values. This work provides basic quantitative information for clinicians to understand the influence of composite layers of bone in relation to mechanical torque resistances during implant insertion in order to obtain desired primary implant stability.
AB - The specific aim of this study was to examine the relative contributions to the implant insertion torque value (ITV) by cortical and trabecular components of an in vitro bone model. Simulated bone blocks of polyurethane were used with 2 densities of foam (0.08 g/cm3 to mimic trabecular bone and 0.64 g/cm3 to mimic cortical bone). We have developed a new platform technology to collect data that enables quantitative evaluation of ITV at different implant locations. Seven groups were used to model varying thicknesses of cortical bone over a lower-quality trabecular bone that have clinical significance: A solid 0.08 g/cm3 block; 1 mm, 2 mm, and 3 mm thick 0.64 g/cm3 sheets with no underlayer; and 1 mm, 2 mm, and 3 mm thick 0.64 g/cm3 sheets laminated on top of a 4 cm thick 0.08 g/cm3 block. The ITVs were recorded as a function of insertion displacement distance. Relative contributions of ITV ranged from 3% to 18% from trabecular bone, and 62% to 74% from cortical bone depending on the thickness of the cortical layer. Inserting an implant into 2-mm and 3-mm cortical layers laminated atop trabecular blocks had a synergistic effect on ITVs. Finally, an implant with a reverse bevel design near the abutment showed final average torque values that were 14% to 34% less than their maximum torque values. This work provides basic quantitative information for clinicians to understand the influence of composite layers of bone in relation to mechanical torque resistances during implant insertion in order to obtain desired primary implant stability.
KW - Implant design
KW - Insertion torque
KW - Mechanical contribution of cortical bone
KW - Primary implant stability
KW - Torque-displacement curve
UR - https://www.scopus.com/pages/publications/84962888211
U2 - 10.1563/aaid-joi-D-14-00322
DO - 10.1563/aaid-joi-D-14-00322
M3 - Article
C2 - 26103153
AN - SCOPUS:84962888211
SN - 0160-6972
VL - 42
SP - 145
EP - 152
JO - Journal of Oral Implantology
JF - Journal of Oral Implantology
IS - 2
ER -