Skip to main navigation Skip to search Skip to main content

Postmortem analysis of anastomotic suture line disruption following carotid endarterectomy

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

The tensile strength of a surgical suture is essential in maintaining the integrity of vascular anastomoses. While faulty operative technique and the loading strength of individual sutures have been implicated in spontaneous suture line disruptions, there has, to date, never been a published postmortem analysis of a suture that has known to have failed. We present the case of suture line disruption leading to fatal exsanguination in a 77-year-old man following carotid endarterectomy with a facial vein patch. Using both dissecting and scanning electron microscopy, we determined that surgical technique (an untied knot) was the cause of the suture line disruption. The removal of a broken or untied suture at surgery or at autopsy should not preclude proper analysis of the failed suture, because the results can have both medico-legal and public health implications.

Original languageEnglish
Pages (from-to)1077-1081
Number of pages5
JournalJournal of Forensic Sciences
Volume49
Issue number5
DOIs
StatePublished - Sep 2004

Keywords

  • Anastomosis
  • Autopsy
  • Carotid endarterectomy
  • Forensic pathology
  • Forensic science
  • Polypropylene
  • Postmortem
  • Suture

Fingerprint

Dive into the research topics of 'Postmortem analysis of anastomotic suture line disruption following carotid endarterectomy'. Together they form a unique fingerprint.

Cite this