Abstract
Background: Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. Case: We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose ProGlide Suture-Mediated Closure was used for closure. The closure was complicated by a separation of the ProGlide device resulting in the migration of the footplate to the descending aorta. Correction required snare retrieval via radial access, and the patient recovered without complications. Discussion: We highlight an important complication of the Perclose ProGlide Suture-Mediated Closure device that is rare but important to know when performing endovascular closures with this device.