T1 - Multicenter clinical and analytical evaluation of the AxSYM troponin-I immunoassay to assist in the diagnosis of myocardial infarction
AU - Apple, Fred S.
AU - Maturen, Andrew J.
AU - Mullins, Richard E.
AU - Painter, Pennell C.
AU - Pessin-Minsley, Melissa S.
AU - Webster, Robert A.
AU - Flores, Jennifer Spray
AU - DeCresce, Robert
AU - Fink, Daniel J.
AU - Buckley, Patrice M.
AU - Marsh, Julie
AU - Ricchiuti, Vincent
AU - Christenson, Robert H.
PY - 1999
Y1 - 1999
N2 - We evaluated the AxSYM troponin I (cTnI) immunoassay for assisting in the detection of acute myocardial infarction (AMI). At four sites, the total imprecision (CV) over 20 days was 6.3-10.2%. The minimum detectable concentration was 0.14 ± 0.05 μg/L. Comparison of cTnI measurements between the AxSYM and Stratus (n = 406) over the dynamic range of the AxSYM assay demonstrated good correlation, r = 0.881, with a proportional bias: AxSYM cTnI = 3.50(Stratus cTnI) - 1.10. The confidence intervals (95%) for the slope and intercept were 3.39-3.64 and -1.32 to -0.95, respectively. The expected cTnI concentration in healthy individuals was ≤0.5 μg/L, whereas the ROC curve-determined cutoff for AMI was 2.0 μg/L. This gave a diagnostic sensitivity of 91.8% and specificity of 92.4% when tested in serial samples collected within 24 h of admission in 633 patients presenting with chest pain, of which 122 had an AMI. The concordances of the AxSYM cTnI with the Stratus cTnI, OPUS cTnI, and Access cTnI were 95.3%, 95.1%, and 94.3%, respectively, from patients with suspected AMI. The AxSYM cTnI demonstrated excellent clinical specificity, ≥96%, in skeletal muscle injury, chronic renal disease, and same-day noncardiac surgery patients.
AB - We evaluated the AxSYM troponin I (cTnI) immunoassay for assisting in the detection of acute myocardial infarction (AMI). At four sites, the total imprecision (CV) over 20 days was 6.3-10.2%. The minimum detectable concentration was 0.14 ± 0.05 μg/L. Comparison of cTnI measurements between the AxSYM and Stratus (n = 406) over the dynamic range of the AxSYM assay demonstrated good correlation, r = 0.881, with a proportional bias: AxSYM cTnI = 3.50(Stratus cTnI) - 1.10. The confidence intervals (95%) for the slope and intercept were 3.39-3.64 and -1.32 to -0.95, respectively. The expected cTnI concentration in healthy individuals was ≤0.5 μg/L, whereas the ROC curve-determined cutoff for AMI was 2.0 μg/L. This gave a diagnostic sensitivity of 91.8% and specificity of 92.4% when tested in serial samples collected within 24 h of admission in 633 patients presenting with chest pain, of which 122 had an AMI. The concordances of the AxSYM cTnI with the Stratus cTnI, OPUS cTnI, and Access cTnI were 95.3%, 95.1%, and 94.3%, respectively, from patients with suspected AMI. The AxSYM cTnI demonstrated excellent clinical specificity, ≥96%, in skeletal muscle injury, chronic renal disease, and same-day noncardiac surgery patients.
UR - https://www.scopus.com/pages/publications/0033035442
U2 - 10.1093/clinchem/45.2.206
DO - 10.1093/clinchem/45.2.206
M3 - Article
C2 - 9931042
AN - SCOPUS:0033035442
SN - 0009-9147
VL - 45
SP - 206
EP - 212
JO - Clinical Chemistry
JF - Clinical Chemistry
IS - 2
ER -