T1 - Management of acute cardiac failure with mechanical assist
T2 - Experience with the ABIOMED BVS 5000
AU - Samuels, Louis E.
AU - Holmes, Elena C.
AU - Thomas, Matthew P.
AU - Entwistle, John C.
AU - Morris, Rohinton J.
AU - Narula, Jaguat
AU - Wechsler, Andrew S.
PY - 2001
Y1 - 2001
N2 - Background. Mechanical circulatory assist industries have developed ventricular assist devices (VAD) for short-, intermediate-, and long-term use. The purpose of this report is to describe the progress made with the ABIOMED Biventricular System (BVS) 5000 (ABIOMED, Inc, Danvers, MA) short-term VAD. Methods. From June 1994 through August 2000, all cardiogenic shock patients who required short-term mechanical assist were supported with the ABIOMED BVS 5000. Insertion criteria included any condition that may potentially result in cardiac recovery. A formal algorithm for timing of insertion was established to standardize implantation criteria. Results. A total of 45 patients were supported at Hahnemann University Hospital, Philadelphia, PA. There were 26 male and 19 female patients, with a mean age of 57.9 years (range 33 to 80 years). Devices were inserted for postcardiotomy shock in 36 patients (80%) and precardiotomy shock in 9 patients (20%). The average duration of support was 8.3 days (range 1 to 31 days). Overall, there were 22 (49%) patients weaned from support and 14 (31%) discharged from the hospital. For patients in whom the device was implanted in accordance with an established protocol (group A), the wean and discharge rates were 60% and 43%, respectively. The most common morbidities included bleeding and adverse neurologic events. Conclusions. The ABIOMED BVS 5000 VAD continues to be a valuable form of short-term mechanical assist for acute cardiogenic shock. The formation of a uniform VAD insertion algorithm has helped to standardize protocols in management.
AB - Background. Mechanical circulatory assist industries have developed ventricular assist devices (VAD) for short-, intermediate-, and long-term use. The purpose of this report is to describe the progress made with the ABIOMED Biventricular System (BVS) 5000 (ABIOMED, Inc, Danvers, MA) short-term VAD. Methods. From June 1994 through August 2000, all cardiogenic shock patients who required short-term mechanical assist were supported with the ABIOMED BVS 5000. Insertion criteria included any condition that may potentially result in cardiac recovery. A formal algorithm for timing of insertion was established to standardize implantation criteria. Results. A total of 45 patients were supported at Hahnemann University Hospital, Philadelphia, PA. There were 26 male and 19 female patients, with a mean age of 57.9 years (range 33 to 80 years). Devices were inserted for postcardiotomy shock in 36 patients (80%) and precardiotomy shock in 9 patients (20%). The average duration of support was 8.3 days (range 1 to 31 days). Overall, there were 22 (49%) patients weaned from support and 14 (31%) discharged from the hospital. For patients in whom the device was implanted in accordance with an established protocol (group A), the wean and discharge rates were 60% and 43%, respectively. The most common morbidities included bleeding and adverse neurologic events. Conclusions. The ABIOMED BVS 5000 VAD continues to be a valuable form of short-term mechanical assist for acute cardiogenic shock. The formation of a uniform VAD insertion algorithm has helped to standardize protocols in management.
UR - http://www.scopus.com/inward/record.url?scp=0035100666&partnerID=8YFLogxK
U2 - 10.1016/S0003-4975(00)02644-8
DO - 10.1016/S0003-4975(00)02644-8
M3 - Article
C2 - 11265869
AN - SCOPUS:0035100666
SN - 0003-4975
VL - 71
SP - S67-S72
JO - Annals of Thoracic Surgery
JF - Annals of Thoracic Surgery
IS - 3 SUPPL.
ER -