TY - JOUR
T1 - Intermediate-term outcomes of aortic valve replacement using a bioprosthesis with a novel tissue
AU - The COMMENCE Trial Investigators
AU - Johnston, Douglas R.
AU - Griffith, Bartley P.
AU - Puskas, John D.
AU - Bavaria, Joseph E.
AU - Svensson, Lars G.
AU - Blackstone, Eugene H.
AU - Gammie, James S.
AU - Heimansohn, David A.
AU - Sadowski, Jerzy
AU - Bartus, Krzysztof
AU - Rozanski, Jacek
AU - Bilewskai, Agata
AU - Rosengart, Todd
AU - Girardi, Leonard N.
AU - Klodell, Charles T.
AU - Mumtaz, Mubashir A.
AU - Takayama, Hiroo
AU - Halkos, Michael
AU - Starnes, Vaughn
AU - Boateng, Percy
AU - Timek, Tomasz A.
AU - Ryan, William
AU - Omer, Shuab
AU - Smith, Craig R.
N1 - Publisher Copyright:
© 2020
PY - 2021/11
Y1 - 2021/11
N2 - Objectives: The COMMENCE trial was conducted to evaluate the safety and effectiveness of an aortic bioprosthesis with novel RESILIA tissue (Edwards Lifesciences, Irvine, Calif). Reports of early noncalcific valve failure resulting from thrombosis or leaflet tears in other valves warrant careful evaluation of early valve performance. Methods: Patients underwent clinically indicated surgical aortic valve replacement with the Edwards Pericardial Aortic Bioprosthesis, Model 11000A (Edwards Lifesciences) in a prospective, multinational, multicenter (n = 27), single-arm, Food and Drug Administration Investigational Device Exemption trial. Events were adjudicated by an independent clinical events committee; echocardiograms were analyzed by an independent core laboratory. Results: Between January 2013 and March 2016, 689 patients received the study valve. Mean age was 67.0 ± 11.6 years. Mean Society of Thoracic Surgeons predicted risk of mortality was 2.0% ± 1.8%. Follow-up duration was 3.7 ± 1.2 years, with a total of 2533 patient years of follow-up and a median follow-up of 4 years. Early all-cause mortality was 1.2%, thromboembolism 2.3%, all bleeding 1.0%, and major paravalvular leak 0.1%. One- and 4-year actuarial freedom from all-cause mortality was 97.7% (95% confidence interval, 96.5%-98.8%) and 91.9% (95% confidence interval, 89.7%-94.1%), respectively. At 4 years, New York Heart Association functional class improved compared with baseline in 63.0%, effective orifice area was 1.5 ± 0.5 cm2, and mean gradient was 11.0 ± 5.6 mm Hg. Freedom from moderate or greater transvalvular insufficiency was 99.7%. There were no events of structural valve deterioration. Conclusions: Safety and hemodynamic performance of this aortic bioprosthesis with RESILIA tissue at 4 years are favorable. This novel tissue does not appear to result in unexpected early thrombosis events or noncalcific structural valve deterioration.
AB - Objectives: The COMMENCE trial was conducted to evaluate the safety and effectiveness of an aortic bioprosthesis with novel RESILIA tissue (Edwards Lifesciences, Irvine, Calif). Reports of early noncalcific valve failure resulting from thrombosis or leaflet tears in other valves warrant careful evaluation of early valve performance. Methods: Patients underwent clinically indicated surgical aortic valve replacement with the Edwards Pericardial Aortic Bioprosthesis, Model 11000A (Edwards Lifesciences) in a prospective, multinational, multicenter (n = 27), single-arm, Food and Drug Administration Investigational Device Exemption trial. Events were adjudicated by an independent clinical events committee; echocardiograms were analyzed by an independent core laboratory. Results: Between January 2013 and March 2016, 689 patients received the study valve. Mean age was 67.0 ± 11.6 years. Mean Society of Thoracic Surgeons predicted risk of mortality was 2.0% ± 1.8%. Follow-up duration was 3.7 ± 1.2 years, with a total of 2533 patient years of follow-up and a median follow-up of 4 years. Early all-cause mortality was 1.2%, thromboembolism 2.3%, all bleeding 1.0%, and major paravalvular leak 0.1%. One- and 4-year actuarial freedom from all-cause mortality was 97.7% (95% confidence interval, 96.5%-98.8%) and 91.9% (95% confidence interval, 89.7%-94.1%), respectively. At 4 years, New York Heart Association functional class improved compared with baseline in 63.0%, effective orifice area was 1.5 ± 0.5 cm2, and mean gradient was 11.0 ± 5.6 mm Hg. Freedom from moderate or greater transvalvular insufficiency was 99.7%. There were no events of structural valve deterioration. Conclusions: Safety and hemodynamic performance of this aortic bioprosthesis with RESILIA tissue at 4 years are favorable. This novel tissue does not appear to result in unexpected early thrombosis events or noncalcific structural valve deterioration.
KW - aortic valve replacement, bioprosthetic valve, RESILIA, tissue
UR - https://www.scopus.com/pages/publications/85081216750
U2 - 10.1016/j.jtcvs.2020.01.095
DO - 10.1016/j.jtcvs.2020.01.095
M3 - Article
C2 - 32340804
AN - SCOPUS:85081216750
SN - 0022-5223
VL - 162
SP - 1478
EP - 1485
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
IS - 5
ER -