High Flux Hemofiltration

Juan P. Bosch, Robert Geronemus, Sheldon Glabman, Michael Lysaght, Thomas Kahn, Beat von Albertini

Research output: Contribution to journalArticlepeer-review

20 Scopus citations

Abstract

Experiments were performed using a new hollow fiber hemofilter. Ultrafiltration rates and whole blood urea clearances were measured in post‐dilution hemofiltration. High ultrafiltration rates were obtained with the new hemofilter. No adverse effects were detected despite filtration fractions above 45%. These experiments suggest that it is possible in post‐dilution hemofiltration to obtain small solute clearances comparable to those of hemodialysis without apparent deleterious effects.

Original languageEnglish
Pages (from-to)339-342
Number of pages4
JournalArtificial Organs
Volume2
Issue number4
DOIs
StatePublished - Nov 1978

Keywords

  • hemodialysis
  • hemofiltration
  • post‐dilution hemofiltration
  • ultrafiltration