HemoCue β-glucose photometer evaluated for use in a neonatal intensive care unit

Research output: Contribution to journal › Article › peer-review

10 Scopus citations

Abstract

We evaluated the HemoCue β-Glucose Photometer system for use in our neonatal intensive care unit by assaying 178 heparinized whole-blood samples obtained by heel stick. The required sample size is 5 μL. Plasma glucose was analyzed by a glucose oxidase/oxygen electrode methodology. Across the glucose range of 1.28-21.87 mmol/L, the regression slope was 0.976 (r = 0.976, Sy|x = 0.475). For samples with hematocrit ≤0.30, the regression slope was 0.981 (r = 0.950, Sy|x = 0.415); for hematocrit of 0.31-0.49, the regression slope was 0.984 (r = 0.972, Sy|x = 0.508); and for hematocrit ≥0.50, the regression slope was 0.959 (r = 0.988, Sy|x = 0.394). Human whole blood, bovine whole blood, and bovine serum-based quality-control materials were studied. Except for assays of the low-concentration human control material, the total CV was <3.5%. The accuracy and precision of the HemoCue system were comparable with those of conventional laboratory instrumentation.

Original languageEnglish
Pages (from-to)2329-2332
Number of pages4
JournalClinical Chemistry
Volume39
Issue number11
StatePublished - 1993

Keywords

  • Critical care medicine
  • Hematocrit
  • Pediatric chemistry
  • Point-of-care testing
  • Reflectance photometry