Abstract
The Goldvalve balloon is the only currently available detachable balloon. We undertook a study to determine safety and imaging artifacts in a MR environment. We found no evidence for heating or deflection of the balloon in a comprehensive series of in vitro experiments at 3 T. MR imaging at field strengths up to 3 T of patients with implanted Goldvalve balloons is safe. Imaging artifacts are minimal.
| Original language | English |
|---|---|
| Pages (from-to) | 409-411 |
| Number of pages | 3 |
| Journal | Neuroradiology |
| Volume | 48 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 2006 |
| Externally published | Yes |