Abstract
PURPOSE: To compare the clinical efficacy of a dual-echo fast spin-echo imaging technique, SHARE (shared-view acquisition using repeated echoes), with conventional long-repetition-time spin-echo imaging. METHODS: Conventional spin-echo and SHARE fast spin-echo MR images of the brain were acquired in 50 randomized patients and interpreted separately in conjunction with the T1-weighted images. All images were reviewed independently by two neuroradiologists who were blinded to the clinical history and previous interpretations. RESULTS: The diagnoses rendered for the spin-echo and SHARE images were concordant in 48 of the 50 subjects (96%) by the first reader and in all 50 cases (100%) by the second reader. SHARE images were acquired in one-fourth of the imaging time yet image contrast, quality, and sensitivity to long T2 lesions were comparable. The SHARE technique was less sensitive to hemorrhagic residua. CONCLUSIONS: SHARE is a viable time-saving alternative to the conventional long-repetition-time pulse sequence. Although SHARE images are not as sensitive to magnetic susceptibility effects, the time saved using this technique could be used to perform a gradient-echo sequence when indicated.
| Original language | English |
|---|---|
| Pages (from-to) | 667-673 |
| Number of pages | 7 |
| Journal | American Journal of Neuroradiology |
| Volume | 15 |
| Issue number | 4 |
| State | Published - 1994 |
| Externally published | Yes |