Evaluation of aortic stenosis by continuous wave Doppler ultrasound

Marvin Berger, Russell L. Berdoff, Peter E. Gallerstein, Emanuel Goldberg

Research output: Contribution to journalArticlepeer-review

98 Scopus citations

Abstract

Twenty-four patients with suspected aortic stenosis (Group I) were evaluated noninvasively by continuous wave Doppler ultrasound before undergoing cardiac catheterization. Twenty normal subjects served as the control group (Group II). Maximal velocity measurements in the ascending aorta ranged from 3.0 to 5.8 m/s (mean 4.34 ± 0.65) in Group I versus 1.0 to 1.6 m/s (mean 1.28 ± 0.16) in Group II (p < 0.001). Using the Bernoulli equation, the peak pressure gradient across the aortic valve was calculated from the maximal velocity in the Group I patients. The results correlated well with the peak aortic valve gradient obtained at cardiac catheterization (r = 0.79). In 20 of these 24 patients, the peak Doppler gradient was within 25% of the gradient found at cardiac catheterization. In three patients, the Doppler study underestimated the gradient by slightly more than 25% but still detected the presence of significant aortic stenosis. The Doppler technique failed to detect critical aortic stenosis in only one patient. Significant overestimation of the gradient by Doppler measurement did not occur in any patient. The technique was particularly helpful in older patients in whom other noninvasive tests often yield inconclusive results. An important but infrequent limitation of the technique is underestimation of the gradient that occurs when the angle of incidence between the ultrasound beam and aortic blood flow is too large. The findings indicate that continuous wave Doppler ultrasound provides a reliable estimate of the valvular gradient in most patients with aortic stenosis.

Original languageEnglish
Pages (from-to)150-156
Number of pages7
JournalJournal of the American College of Cardiology
Volume3
Issue number1
DOIs
StatePublished - 1984
Externally publishedYes

Fingerprint

Dive into the research topics of 'Evaluation of aortic stenosis by continuous wave Doppler ultrasound'. Together they form a unique fingerprint.

Cite this