Effects of skin surface temperature on photoplethysmograph

In Cheol Jeong, Hyungro Yoon, Hyunjeong Kang, Hojun Yeom

Research output: Contribution to journalArticlepeer-review

24 Scopus citations

Abstract

Photoplethysmograph (PPG) has been widely used to investigate various cardiovascular conditions. Previous studies demonstrated effects of temperature of the measurement environment; however, an integrated evaluation has not been established in environments with gradual air temperature variation. The purpose of this study is to investigate variations and relationships of blood pressure (BP), PPG and cardiovascular parameters such as heart rate (HR), stroke volume (SV), cardiac output (CO) and total peripheral resistance (TPR), by changing skin surface temperature (SST). Local mild cooling and heating was conducted on 16 healthy subjects. The results showed that local SST changes affected Finometer blood pressures (Finger BP), PPG components and TPR, but not the oscillometric blood pressure (Central BP), HR, SV and CO, and indicated that temperature must be maintained and monitored to reliably evaluate cardiovascular conditions in temperature-varying environments.

Original languageEnglish
Pages (from-to)429-438
Number of pages10
JournalJournal of Healthcare Engineering
Volume5
Issue number4
DOIs
StatePublished - 1 Dec 2014
Externally publishedYes

Keywords

  • Cardiovascular status
  • Local cooling and heating
  • Photoplethysmography
  • Skin surface temperature
  • Vasoconstriction
  • Vasodilation

Fingerprint

Dive into the research topics of 'Effects of skin surface temperature on photoplethysmograph'. Together they form a unique fingerprint.

Cite this