Direct Aspiration Catheter Fracture and Retrieval during Neurothrombectomy

Pouria Moshayedi, Ashutosh P. Jadhav

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

Background: Application of direct aspiration catheters has revolutionized acute stroke care and has led to significant improvement in clinical outcome with a good safety profile. Catheter fracture and retention is a rare but potentially devastating complication. Case Description: Here we present two cases of acute stroke complicated by aspiration catheter fracture and retention. Successful catheter retrieval and revascularization was achieved in both cases. The stenosis or tortuosity of vascular anatomy appears to be the probable contributor to catheter breakage by anchoring the catheter with resultant fracture at the constraint point from catheter withdrawal tensile stress. Conclusion: This report describes application of snare devices in retrieving a broken catheter during thrombectomy in the anterior and posterior circulation, and therefore presents a technique that can be safely utilized to address catheter breakage complicating thrombectomy in different vascular anatomic locations.

Original languageEnglish
Pages (from-to)148-152
Number of pages5
JournalInterventional Neurology
Volume7
Issue number3-4
DOIs
StatePublished - 1 Apr 2018
Externally publishedYes

Keywords

  • Catheter breakage
  • Complication
  • Retrieval
  • Snare devices
  • Thrombectomy

Fingerprint

Dive into the research topics of 'Direct Aspiration Catheter Fracture and Retrieval during Neurothrombectomy'. Together they form a unique fingerprint.

Cite this