Skip to main navigation Skip to search Skip to main content

Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients

  • R. Larbig
  • , L. J. Motloch
  • , M. Bettin
  • , A. Fischer
  • , N. Bode
  • , G. Frommeyer
  • , A. Loeher
  • , J. Koebe
  • , F. Reinke
  • , L. Eckardt

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Aims: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients. Background: TWOS is a feared complication after implantation. It may lead to harmful IS. To date, specific strategies to reduce these events are lacking. Methods: In this retrospective single-centre trial we analysed 146 S‑ICD patients who were implanted between 2010 and 2016. In all eligible consecutive patients (n = 139), follow-up of at least 6 weeks was studied. The incidence of TWOS/IS was analysed in patients receiving a 2nd generation S‑ICD (Emblem-S-ICD) between 2014 and 2016 (Emblem). Their outcome was compared with a control group (SQ) treated with the SQ1010 device between 2010 and 2014, who were followed up for a maximum of 2 years. Furthermore, to test if the software update SMR8 reduces inappropriate shocks in the SQ1010-S-ICD population, the incidence of TWOS/IS was evaluated before and after update installation. Results: Basic characteristics and indications for S‑ICD implantation were similar in both groups. However, the cumulative incidence of TWOS/IS was significantly decreased in Emblem vs. SQ (SQ: 15.4%, n = 14/91 vs. Emblem 4.2%, n = 2/48; p = 0.049). Furthermore, with regards to the SQ population we also observed a trend towards a significant reduction of TWOS/IS after installation of the software update SMR8 in 2014 (before update: 13.4%, n = 11/82 vs. after update: 4.6%, 3/65, p = 0.07). Conclusion: 2nd generation devices but probably also the SMR8 software update reduce the incidence of TWOS/IS in S‑ICD patients.

Original languageEnglish
Pages (from-to)606-611
Number of pages6
JournalNetherlands Heart Journal
Volume26
Issue number12
DOIs
StatePublished - 1 Dec 2018
Externally publishedYes

Keywords

  • Inappropriate shocks
  • Oversensing
  • S-ICD

Fingerprint

Dive into the research topics of 'Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients'. Together they form a unique fingerprint.

Cite this