Skip to main navigation Skip to search Skip to main content

Cross reactivity of the EMIT digoxin assay with digoxin metabolites, and validation of the method for measurement of urinary digoxin

Research output: Contribution to journal › Article › peer-review

6 Scopus citations

Abstract

In evaluating the EMIT (Syva Co.) digoxin assay, we found no cross reactivity between dihydrodigoxin, the major digoxin metabolite, and the EMIT digoxin antibody from two different lots. However, the antibody does cross react, essentially completely, with the digoxin hydrolysis metabolites digoxigenin, digoxigenin mono-digitoxide, and digoxigenin bis-digitoxide. The EMIT method can be used to measure digoxin in urine diluted at least 50- to 100-fold with digoxin-free human plasma; the inter-assay coefficient of variation of this assay is 6%. In addition, we validated the manual EMIT serum digoxin assay, using external ('TRI-rac') quality controls.

Original languageEnglish
Pages (from-to)175-177
Number of pages3
JournalClinical Chemistry
Volume29
Issue number1
DOIs
StatePublished - 1983
Externally publishedYes

Fingerprint

Dive into the research topics of 'Cross reactivity of the EMIT digoxin assay with digoxin metabolites, and validation of the method for measurement of urinary digoxin'. Together they form a unique fingerprint.

Cite this