Skip to main navigation Skip to search Skip to main content

Calculating the uncertainty in lead concentration for in vivo bone lead x-ray fluorescence

Research output: Contribution to journalArticlepeer-review

19 Scopus citations

Abstract

A revised mathematical treatment of the calibration line intercept has been published for in vivo bone lead measurements using 109Cd-based K-shell x-ray fluorescence. The revised calibration line treatment prompts changes, presented herein, to the method for calculating the measurement uncertainty.

Original languageEnglish
Pages (from-to)2033-2039
Number of pages7
JournalPhysics in Medicine and Biology
Volume48
Issue number13
DOIs
StatePublished - 7 Jul 2003

Fingerprint

Dive into the research topics of 'Calculating the uncertainty in lead concentration for in vivo bone lead x-ray fluorescence'. Together they form a unique fingerprint.