Skip to main navigation Skip to search Skip to main content

Calculating bone-lead measurement variance

Research output: Contribution to journalArticlepeer-review

30 Scopus citations

Abstract

The technique of 109Cd-based X-ray fluorescence (XRF) measurements of lead in bone is well established. A paper by some XRF researchers [Gordon CL, et al. The Reproducibility of 109Cd-based X-ray Fluorescence Measurements of Bone Lead. Environ Health Perspect 102:690-694 (1994)] presented the currently practiced method for calculating the variance of an vivo measurement once a calibration line has been established. This paper corrects typographical errors in the method published by those authors; presents a crude estimate of the measurement error that can be acquired without computational peak fitting programs; and draws attention to the measurement error attributable to covariance, an important feature in the construct of the currently accepted method that is flawed under certain circumstances.

Original languageEnglish
Pages (from-to)383-386
Number of pages4
JournalEnvironmental Health Perspectives
Volume108
Issue number5
DOIs
StatePublished - 2000

Keywords

  • Bone
  • Lead
  • Measurement error
  • X-ray fluorescence

Fingerprint

Dive into the research topics of 'Calculating bone-lead measurement variance'. Together they form a unique fingerprint.

Cite this