Abstract
Using a special imaging technique, cartridges were analyzed during a simulated in vitro circulation at 250 and 350 ml/min of blood flow and 25% and 40% hematocrit. Cartridges were placed in vertical position and a cross longitudinal section 1 cm thick was analyzed in sequence by a helical scanner. Dye was injected in the arterial inlet and the progressive distribution was evaluated by sequential densitometrical measures carried out automatically by the machine. The sequential images analyzed by the scanner demonstrated excellent distribution of the flow in the blood compartment with a minimal difference between the central and the peripheral region of the compartment.