Battery longevity of a helix-fixation dual-chamber leadless pacemaker: results from the AVEIR DR i2i Study

  • Vivek Y. Reddy
  • , Rahul Doshi
  • , James E. Ip
  • , Pascal Defaye
  • , Derek V. Exner
  • , Robert Canby
  • , Morio Shoda
  • , Maria G. Bongiorni
  • , Gerhard Hindricks
  • , Petr Neužil
  • , Thomas Callahan
  • , Nima Badie
  • , Matthew G. Fishler
  • , Reinoud E. Knops

Research output: Contribution to journal › Article › peer-review

23 Scopus citations

Abstract

Aims A dual-chamber leadless pacemaker (LP) system that employs distinct atrial and ventricular LP devices (ALP, VLP) has been introduced to clinical practice. Proprietary, low-energy, implant-to-implant (i2i) communication at each beat enables the devices to maintain synchronous atrioventricular sensing and pacing. We evaluated device longevities and contributing factors, such as i2i communication. Methods Patients meeting dual-chamber pacing indications received the dual-chamber LP system as part of a prospective, multi- and results centre, international clinical trial (Aveir DR i2i Study, NCT05252702). Programming and diagnostics were interrogated from all de novo, non-revised, dual-chamber programmed devices at 12 months post-implant. This analysis included 302 patients (65% male; age 70 ± 13 years; weight 80 ± 19 kg; intrinsic heart rate 55 ± 7 bpm; 58% sinus node dysfunction, 27% atrioventricular block). At 12 months, devices were programmed to dual-chamber pacing (DDD(R) or DDI(R)) at a median 60 bpm rate, median 1.25 V pulse amplitude in ALP and 1.5 V in VLP, median 0.4 ms pulse width, and median i2i signal setting level 5 out of 7. Median ALP and VLP remaining battery longevities at 12 months were 4.3 and 9.1 years, with median total ALP and VLP longevities of 5.3 and 9.9 years. Base rate, pulse amplitude, pacing percentage, event rate, impedance, and i2i setting level all exhibited significant correlations with ALP and VLP longevities (P < 0.001). Programming i2i setting levels below 7 produced the greatest longevity savings. Conclusion The first dual-chamber LP demonstrated adequate projected battery longevity after 12 months of use. Patient-specific device programming considerations, unique to leadless devices, may extend longevity (Figure presented).

Original languageEnglish
Article numbereuaf074
JournalEuropace
Volume27
Issue number6
DOIs
StatePublished - 1 Jun 2025

Keywords

  • Aveir
  • Battery longevity
  • Dual-chamber
  • Leadless pacemaker

Fingerprint

Dive into the research topics of 'Battery longevity of a helix-fixation dual-chamber leadless pacemaker: results from the AVEIR DR i2i Study'. Together they form a unique fingerprint.

Cite this