TY - JOUR
T1 - Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter
T2 - a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardium
AU - Zito, Elio
AU - Mansour, Moussa
AU - Reddy, Vivek Y.
AU - Al-Ahmad, Amin
AU - La Fazia, Vincenzo Mirco
AU - Gianni, Carola
AU - David Burkhardt, J.
AU - Mohanty, Sanghamitra
AU - Papaioannou, Thanassis
AU - Sharma, Tushar
AU - Di Biase, Luigi
AU - Natale, Andrea
N1 - Publisher Copyright:
© 2025 The Author(s).
PY - 2025/11/1
Y1 - 2025/11/1
N2 - Aims Pulsed Field ablation (PFA) is a non-thermal ablation modality with functional myocardial sparing. Recent evidence suggests that clinically used PFA systems may produce non-negligible thermal effects, particularly at the tissue-catheter interface, potentially increasing the risk of thrombo-embolic complications. This study sought to characterize the temperature dynamics of two PFA pulse waveforms, delivered using a variable-loop circular catheter, under different irrigation conditions. Methods and results In total, 132 ablations were performed on 31 bovine myocardial tissue specimens, using Sequence_1 and Sequence_2 at two different irrigation rates - 4 and 30 mL/min. Maximum temperatures and their rise from baseline were measured at the tissue surface, and at 3 and 7 mm depths, across different ablation conditions. Sequence_1 at 4 mL/min produced the highest surface temperature (56.4°C [54.9-58.4]) and temperature rise (ΔT: 19.4°C [17.9-21.4]). Sequence_2 at 30 mL/min showed the most favourable thermal profile, with significantly lower surface temperatures (40.8°C [37.9-43.0], ΔT: 3.8°C [0.9-6.0], P < 0.0001). At 3 mm depth, temperature increases were reduced for all settings, but remained highest with Sequence_1 at 4 mL/min (42.5°C, ΔT: 5.5°C). At 7 mm depth, temperatures remained close to baseline. Both waveform and irrigation optimization independently and synergistically reduced tissue heating, with the Sequence_2 at 30 mL/min achieving the lowest thermal load at all depths. Conclusion This study confirms that PFA can induce relevant thermal effects, especially at the tissue interface. However, waveform optimization and active cooling significantly mitigate these effects. Such strategies to minimize thermal effects should be implemented in clinical practice to enhance procedural safety.
AB - Aims Pulsed Field ablation (PFA) is a non-thermal ablation modality with functional myocardial sparing. Recent evidence suggests that clinically used PFA systems may produce non-negligible thermal effects, particularly at the tissue-catheter interface, potentially increasing the risk of thrombo-embolic complications. This study sought to characterize the temperature dynamics of two PFA pulse waveforms, delivered using a variable-loop circular catheter, under different irrigation conditions. Methods and results In total, 132 ablations were performed on 31 bovine myocardial tissue specimens, using Sequence_1 and Sequence_2 at two different irrigation rates - 4 and 30 mL/min. Maximum temperatures and their rise from baseline were measured at the tissue surface, and at 3 and 7 mm depths, across different ablation conditions. Sequence_1 at 4 mL/min produced the highest surface temperature (56.4°C [54.9-58.4]) and temperature rise (ΔT: 19.4°C [17.9-21.4]). Sequence_2 at 30 mL/min showed the most favourable thermal profile, with significantly lower surface temperatures (40.8°C [37.9-43.0], ΔT: 3.8°C [0.9-6.0], P < 0.0001). At 3 mm depth, temperature increases were reduced for all settings, but remained highest with Sequence_1 at 4 mL/min (42.5°C, ΔT: 5.5°C). At 7 mm depth, temperatures remained close to baseline. Both waveform and irrigation optimization independently and synergistically reduced tissue heating, with the Sequence_2 at 30 mL/min achieving the lowest thermal load at all depths. Conclusion This study confirms that PFA can induce relevant thermal effects, especially at the tissue interface. However, waveform optimization and active cooling significantly mitigate these effects. Such strategies to minimize thermal effects should be implemented in clinical practice to enhance procedural safety.
KW - Atrial fibrillation
KW - Catheter ablation
KW - Heating
KW - Pulse field ablation
KW - VLCC
UR - https://www.scopus.com/pages/publications/105022272117
U2 - 10.1093/europace/euaf278
DO - 10.1093/europace/euaf278
M3 - Article
C2 - 41166649
AN - SCOPUS:105022272117
SN - 1099-5129
VL - 27
JO - Europace
JF - Europace
IS - 11
M1 - euaf278
ER -