Abstract
To examine the fidelity of ink color identification using light microscopy (LM), telemicroscopy (TM), and virtual microscopy (VM). Methods: Twenty H and E-stained frozen section slides, prepared after tissue inking with five stain combinations, were assessed by three pathologists using LM, TM, and VM. TM was performed using Mikroscan D2 slide scanner/LiveQ software with various objectives. VM was performed using Mikroscan D2 scanner/Qumulus software, specimens digitized at 20. Results: Sensitivity/specificity by LM was 100%/100% for all colors. TM showed high overall specificity but poor sensitivity, particularly red (54%). VM showed high specificity for all colors except black (69%) and, consequently, poor sensitivity for all colors except black (96%).
| Original language | English |
|---|---|
| Pages (from-to) | 95-99 |
| Number of pages | 5 |
| Journal | American Journal of Clinical Pathology |
| Volume | 151 |
| Issue number | 1 |
| DOIs | |
| State | Published - 1 Jan 2019 |
| Externally published | Yes |
Keywords
- Telemicroscopy
- Telepathology
- Tissue-marking dye
- Virtual microscopy