Abstract
An improved method for the quantitation of 3-methoxy-4-hydroxyphenylethyleneglycol in cerebrospinal fluid is described. Sample cleaning was done by SEP PAK C18 Cartridge prior to the MHPG assay by high-performance liquid chromatography with electrochemical detector. The results are in good agreement wilth the GC/MS method. The average recovery is 68.3+4.1%, within run coefficient of variation of 3.8% and day to day of 7.2%. The method is simple, sensitive and accurate and can be used for routine work.
| Original language | English |
|---|---|
| Pages (from-to) | 1997-2003 |
| Number of pages | 7 |
| Journal | Journal of Liquid Chromatography |
| Volume | 6 |
| Issue number | 11 |
| DOIs | |
| State | Published - 1 Sep 1983 |