TY - JOUR
T1 - A Multicenter Evaluation of a Closed-Loop Anesthesia Delivery System
T2 - A Randomized Controlled Trial
AU - Puri, Goverdhan D.
AU - Mathew, Preethy J.
AU - Biswas, Indranil
AU - Dutta, Amitabh
AU - Sood, Jayashree
AU - Gombar, Satinder
AU - Palta, Sanjeev
AU - Tsering, Morup
AU - Gautam, P. L.
AU - Jayant, Aveek
AU - Arora, Inderjeet
AU - Bajaj, Vishal
AU - Punia, T. S.
AU - Singh, Gurjit
N1 - Publisher Copyright:
© 2015 International Anesthesia Research Society.
PY - 2016/1/1
Y1 - 2016/1/1
N2 - BACKGROUND: Closed-loop systems for anesthesia delivery have been shown to outperform traditional manual control in different clinical settings. The present trial was aimed at evaluating the feasibility and efficacy of Bispectral Index (BISTM)-guided closed-loop anesthesia delivery system (CLADS) in comparison with manual control across multiple centers in India. METHODS: Adult patients scheduled for major surgical procedures of an expected duration of 1 to 3 hours were randomized across 6 sites into 2 groups: a CLADS group and a manual group. In the manual control group, propofol infusion was titrated manually by the attending anesthesiologist to a BIS of 50 during induction and maintenance. Analgesia was maintained with fentanyl infusion and nitrous oxide in both groups. In the CLADS group, both induction and maintenance of anesthesia were performed automatically using CLADS. The primary outcome measure was the performance of the system as assessed by the percentage of total anesthesia time BIS remained ±10 of target BIS. The secondary outcome measures were a percentage of anesthesia-time heart rate and mean arterial pressure within 25% of the baseline, median absolute performance error, wobble, and global score. Wobble indicates intraindividual variability in the control of BIS, and global score reflects the overall performance; lower values indicate superior performance for both parameters. The performance parameters of the system also were compared among the participating sites. RESULTS: Two hundred forty-two patients were randomized. BIS was maintained within ±10 of target for significantly longer time in the CLADS group (81.4% ± 8.9 % of anesthesia duration) than in the manual group (55.34% ± 25%, P < 0.0001). The indices that assess performance were significantly better in the CLADS group than the manual group as follows: median absolute performance error was 10 (10, 12) (median [interquartile range]) in the CLADS group versus 18 (14, 24) in the manual group, P < 0.0001; wobble was 9 (8, 10) in CLADS group versus 10 (8, 14) in the manual group, P = 0.0009; and Global score, which reflects overall performance, was 24 (19, 30) in the CLADS group versus 51 (31, 99) in the manual group, P < 0.0001. The percentage of time heart rate was within 25% of the baseline was significantly greater in the CLADS group (heart rate of 95 [87, 99], median [interquartile range], in the CLADS group versus 90 [75, 98] in the manual group P = 0.0031). On comparison of data between the centers, the performance parameters did not differ significantly among the centers in the CLADS group (P = 0.94), but the parameters differed significantly among the centers in the manual group (P < 0.001). CONCLUSIONS: Our study in a multicenter setting proves the consistently better performance of automated anesthesia drug delivery compared with conventional manual control. This highlights an important advantage of an automated system for delivering standardized anesthesia, thereby overcoming differences in practices among anesthesiologists.
AB - BACKGROUND: Closed-loop systems for anesthesia delivery have been shown to outperform traditional manual control in different clinical settings. The present trial was aimed at evaluating the feasibility and efficacy of Bispectral Index (BISTM)-guided closed-loop anesthesia delivery system (CLADS) in comparison with manual control across multiple centers in India. METHODS: Adult patients scheduled for major surgical procedures of an expected duration of 1 to 3 hours were randomized across 6 sites into 2 groups: a CLADS group and a manual group. In the manual control group, propofol infusion was titrated manually by the attending anesthesiologist to a BIS of 50 during induction and maintenance. Analgesia was maintained with fentanyl infusion and nitrous oxide in both groups. In the CLADS group, both induction and maintenance of anesthesia were performed automatically using CLADS. The primary outcome measure was the performance of the system as assessed by the percentage of total anesthesia time BIS remained ±10 of target BIS. The secondary outcome measures were a percentage of anesthesia-time heart rate and mean arterial pressure within 25% of the baseline, median absolute performance error, wobble, and global score. Wobble indicates intraindividual variability in the control of BIS, and global score reflects the overall performance; lower values indicate superior performance for both parameters. The performance parameters of the system also were compared among the participating sites. RESULTS: Two hundred forty-two patients were randomized. BIS was maintained within ±10 of target for significantly longer time in the CLADS group (81.4% ± 8.9 % of anesthesia duration) than in the manual group (55.34% ± 25%, P < 0.0001). The indices that assess performance were significantly better in the CLADS group than the manual group as follows: median absolute performance error was 10 (10, 12) (median [interquartile range]) in the CLADS group versus 18 (14, 24) in the manual group, P < 0.0001; wobble was 9 (8, 10) in CLADS group versus 10 (8, 14) in the manual group, P = 0.0009; and Global score, which reflects overall performance, was 24 (19, 30) in the CLADS group versus 51 (31, 99) in the manual group, P < 0.0001. The percentage of time heart rate was within 25% of the baseline was significantly greater in the CLADS group (heart rate of 95 [87, 99], median [interquartile range], in the CLADS group versus 90 [75, 98] in the manual group P = 0.0031). On comparison of data between the centers, the performance parameters did not differ significantly among the centers in the CLADS group (P = 0.94), but the parameters differed significantly among the centers in the manual group (P < 0.001). CONCLUSIONS: Our study in a multicenter setting proves the consistently better performance of automated anesthesia drug delivery compared with conventional manual control. This highlights an important advantage of an automated system for delivering standardized anesthesia, thereby overcoming differences in practices among anesthesiologists.
UR - http://www.scopus.com/inward/record.url?scp=84950349743&partnerID=8YFLogxK
U2 - 10.1213/ANE.0000000000000769
DO - 10.1213/ANE.0000000000000769
M3 - Article
C2 - 25902324
AN - SCOPUS:84950349743
SN - 0003-2999
VL - 122
SP - 106
EP - 114
JO - Anesthesia and Analgesia
JF - Anesthesia and Analgesia
IS - 1
ER -